Self-Expandable Transcatheter Aortic Valve Frame Infolding: An Increasingly Recognized Complication
Gurpreet Singh1, Vien Le1, Robert J Wiechmann2, Steven W Schreiter1
1 Department of Cardiovascular Medicine, Mayo Clinic Health System, Eau Claire, WI, USA
2 Department of Cardiovascular Surgery, Mayo Clinic Health System, Eau Claire, WI, USA
Doi: 10.12890/2020_002100 - European Journal of Case Reports in Internal Medicine - © EFIM 2020
Received: 25/10/2020
Accepted: 14/11/2020
Published: 14/12/2020

How to cite this article: Singh G, Le V, Wiechmann RJ, Schreiter SW. Self-expandable transcatheter aortic valve frame infolding: an increasingly recognized complication. EJCRIM 2020;7 doi:10.12890/2020_002100.

Conflicts of Interests: The authors declare there are no competing interests.
This article is licensed under a Commons Attribution Non-Commercial 4.0 License

ABSTRACT

Transcatheter aortic valve replacement (TAVR) with either a balloon-expandable or a self-expandable transcatheter heart valve (THV) is an approved therapy for patients with symptomatic severe aortic stenosis and high or intermediate surgical risk. Here we present a case of severe valve frame infolding of a CoreValve Evolut PRO® self-expandable THV (Medtronic Inc.), which was restored to optimal geometry with balloon post-dilation.

LEARNING POINTS

KEYWORDS

Transcatheter aortic valve replacement, cardiac CT scan

CASE DESCRIPTION

A 93-year-old man with symptomatic severe aortic stenosis (AS) was referred to our centre for transcatheter aortic valve replacement (TAVR). Pre-procedural computed tomography (Fig. 1) showed severe trileaflet aortic valve calcification, and extensive calcium extending from the aortic annulus to the left ventricular outflow tract (LVOT). Our heart team decided to proceed with a CoreValve Evolut PRO® self-expandable TAVR (Medtronic Inc. Minneapolis, MN, USA).
The procedure was done under general anaesthesia, and left femoral access was used to advance the equipment. Balloon valvuloplasty was performed with a 20 mm Z-MED balloon (Braun Medical), and an Evolute PRO® valve was then advanced under fluoroscopic guidance. During the procedure, for better positioning we had to recapture the valve twice. After the first recapture, the valve struts looked malaligned. We expected this abnormality to get better after complete deployment, but that did not happen (Fig. 2, left image). A post-procedural echocardiogram and non-contrast ECG-gated CT confirmed severe periprosthetic regurgitation, a high prosthetic gradient (systolic mean Doppler gradient 30 mmHg), and severe infolding of the CoreValve stent frame from the inflow to the waist (Fig. 3, left image; Fig. 4, left image). Our heart team decided to take the patient back to the hybrid operating room for redo balloon valvuloplasty in an attempt to restore the frame geometry and improve transcatheter heart valve (THV) function. We used an Edwards 23 mm balloon catheter, which restored the optimal geometry of the CoreValve frame (Fig. 2, right image; Fig. 4, right image) and significantly improved the periprosthetic regurgitation from severe to mild (Fig. 3, right image) and systolic mean Doppler gradient from 30 to 11 mmHg.

Figure 1 (click to enlarge)

Figure 2 (click to enlarge)

Figure 1. Aortic annulus with eccentric calcification
Figure 2. Fluoroscopy images. Black arrows indicate infolding of the transcatheter valve (left image), which resolved after ballooning of the valve (right image)

Figure 3 (click to enlarge)

Figure 4 (click to enlarge)

Figure 3. Echo images showing improvement in aortic regurgitation (before valve ballooning in the left image, and after ballooning in the right image)
Figure 4. CT scan showing folding of the transcatheter valve in the left image, which resolved after ballooning in the right image

DISCUSSION

The distortion of the valve frame was the result of eccentric calcification causing asymmetric recapturing, which later on resulted in infolding after subsequent deployment. This issue can arise due to eccentric calcium, valve oversizing, and improper loading of the valve [1, 2]. It can also occur during recapture when trying to deploy the valve in a better position as demonstrated by this case. Our case highlights that infolding in a self-expanding valve is unlikely to resolve on its own considering that we waited for few days to allow it to self-correct.
We suggest that after a recapture, the CoreValve should be carefully examined to rule out any distortions. The valve should be replaced with a new one in case of asymmetric recapture, because the valve may or may not be able to expand fully after release. However, if the valve has been deployed, as demonstrated by this case, balloon valvuloplasty with an appropriately sized balloon should be considered. The valve should be crossed with a pigtail to avoid paravalvular crossing or through the valve struts.
There is a theoretical possibility of deploying the valve inside the valve to increase the outward radial force, but it is difficult to comment on the utility of this approach considering the lack of literature.